PRODUCTS

Catalog Number:FS1577H
FS-Human IL-2Rα(Interleukin 2 Receptor Alpha) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human IL-2Rα(Interleukin 2 Receptor Alpha) ELISA Kit Species Human
Uniprot ID P01589 Alternative Names CD25 antigen, CD25, IDDM10, IL-2 R alpha, IL-2 receptor subunit alpha, IL2R alpha, IL-2R subunit alpha, IL2R, IL2RA, IL-2Ra, IL-2-RA, IL2-RA, interleukin 2 receptor, alpha, interleukin-2 receptor subunit alpha, p55, TAC antigen, TCGFR
Detection method Sandwich Sensitivity 1.26 pg/mL
Standard 1000pg/mL Detection Range 15.63-1000pg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids;Sample Volume=50μL
Reaction time 1.5H Research Area Cancer, Immunology
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human IL-2Rα. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human IL-2Rα are added to the micro ELISA plate wells. Human IL-2Rα in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human IL-2Rα. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human IL-2Rα. The concentration of Human IL-2Rα in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
1000.00 2.555 2.497
500.00 1.646 1.588
250.00 0.962 0.904
125.00 0.517 0.459
62.50 0.272 0.214
31.25 0.146 0.088
15.63 0.104 0.046
0.00 0.058
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 58.26 235.98 288.08 58.26 226.54 282.32
Standard deviation 3.35 12.93 12.73 3.47 11.10 10.33
C V (%) 5.75 5.48 4.42 5.96 4.90 3.66
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 88-103 95
EDTA plasma (n=8) 80-95 86
Cell culture media (n=8) 92-106 99
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-102 98-106 89-97
Average (%) 90 105 93
1:4 Range (%) 83-96 98-106 85-94
Average (%) 86 101 87
1:8 Range (%) 87-98 86-99 87-96
Average (%) 95 87 94
1:16 Range (%) 82-95 80-95 92-101
Average (%) 90 80 97
Assay Procedures