PRODUCTS

Catalog Number:FS2216H
FS-Human CYFRA21-1(Cytokeratin Fragment Antigen 21-1) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human CYFRA21-1(Cytokeratin Fragment Antigen 21-1) ELISA Kit Species Human
Uniprot ID P08727 Alternative Names KRT19, CK19, K19, K1CS, keratin 19
Detection method Sandwich Sensitivity 0.15 ng/mL
Standard 40ng/mL Detection Range 0.625-40ng/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids;Sample Volume=50μL
Reaction time 1.5H Research Area Cancer, Immunology;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human CYFRA21-1. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human CYFRA21-1 are added to the micro ELISA plate wells. Human CYFRA21-1 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human CYFRA21-1. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human CYFRA21-1. The concentration of Human CYFRA21-1 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
40.00 2.375 2.313
20.00 1.785 1.723
10.00 0.977 0.915
5.00 0.476 0.414
2.50 0.272 0.210
1.25 0.141 0.079
0.63 0.102 0.040
0.00 0.062
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 2.34 8.90 11.83 2.41 8.81 11.24
Standard deviation 0.16 0.45 0.84 0.13 0.51 0.90
C V (%) 6.84 5.06 7.10 5.39 5.79 8.01
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-92 86
EDTA plasma (n=8) 80-96 88
Cell culture media (n=8) 80-95 86
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-97 85-94 81-93
Average (%) 91 86 87
1:4 Range (%) 86-92 85-94 93-102
Average (%) 90 93 95
1:8 Range (%) 95-103 87-96 80-92
Average (%) 101 89 82
1:16 Range (%) 98-106 82-95 87-98
Average (%) 101 94 95
Assay Procedures