PRODUCTS

Catalog Number:FS2840H
FS-Human NGAL(Neutrophil Gelatinase Associated Lipocalin) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human NGAL(Neutrophil Gelatinase Associated Lipocalin) ELISA Kit Species Human
Uniprot ID P80188 Alternative Names 24p3, 25 kDa alpha-2-microglobulin-related subunit of MMP-9, HNL, LCN2, lipocalin 2 (oncogene 24p3), lipocalin 2, Lipocalin2, Lipocalin-2, migration-stimulating factor inhibitor, MSFI, neutrophil gelatinase-associated lipocalin, NGAL, NGALlipocalin-2, Oncogene 24p3, p25, Siderocalin, Uterocalin
Detection method Sandwich Sensitivity 0.16 ng/mL
Standard 20ng/mL Detection Range 0.31-20ng/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids;Sample Volume=50μL
Reaction time 1.5H Research Area Signaling transduction, Metabolism ;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human NGAL. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human NGAL are added to the micro ELISA plate wells. Human NGAL in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human NGAL. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human NGAL. The concentration of Human NGAL in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
20.00 2.530 2.469
10.00 1.731 1.670
5.00 1.007 0.946
2.50 0.484 0.423
1.25 0.267 0.206
0.63 0.138 0.077
0.32 0.096 0.035
0.00 0.061
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 1.24 3.56 9.38 1.29 3.70 9.00
Standard deviation 0.06 0.28 0.60 0.10 0.21 0.59
C V (%) 4.84 7.87 6.40 7.75 5.68 6.56
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 92-105 98
EDTA plasma (n=8) 84-96 90
Cell culture media (n=8) 80-95 87
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-93 93-102 91-104
Average (%) 91 98 92
1:4 Range (%) 91-105 93-102 89-97
Average (%) 96 101 93
1:8 Range (%) 87-98 93-106 93-102
Average (%) 90 102 97
1:16 Range (%) 89-94 87-101 81-96
Average (%) 94 90 88
Assay Procedures