Product Name GSSG(Oxidized Glutathione) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names GSSG
Detection method Competitive Sensitivity 121.88 ng/mL
Standard 20000ng/mL Detection Range 312.5-20000ng/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids, Sample volume: 50μL
Reaction time 1.5H Research Area Metabolic pathway; Signal Transduction; Cancer, Nutrition metabolism;Hormone metabolism; Signal transduction;Metabolic pathway; Enzyme & Kinase; Signal transduction;Metabolic pathway; Nutrition metabolism;Hormone metabolism;Cardiovascular,Signaling transduction,Neuroscience,Cancer,Metabolism; Enzyme & Kinase;Infection immunity;Endocrinology;Hormone metabolism; Metabolic pathway; Signal Transduction; Cancer
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with GSSG. During the reaction, GSSG in the sample or standard competes with a fixed amount of GSSG on the solid phase supporter for sites on the Biotinylated Detection Ab specific to GSSG. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of GSSG in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
20000.00 0.195
10000.00 0.331
5000.00 0.546
2500.00 0.880
1250.00 1.308
625.00 1.700
312.50 2.076
0.00 2.516
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 393.70 840.34 5617.98 409.84 628.93 6993.01
Standard deviation 41.95 92.71 290.18 52.22 115.40 361.21
C V (%) 5.59 5.81 3.60 6.11 5.77 2.93
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-92 86
EDTA plasma (n=8) 90-105 97
Cell culture media (n=8) 88-102 95
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-94 86-97 87-98
Average (%) 91 94 93
1:4 Range (%) 87-99 86-97 85-92
Average (%) 87 90 86
1:8 Range (%) 85-101 88-95 80-96
Average (%) 96 93 91
1:16 Range (%) 80-90 92-101 89-100
Average (%) 89 101 97