Product Name Human LECT2(Leukocyte cell-derived chemotaxin-2)ELISA Kit Species Human
Uniprot ID O14960 Alternative Names Leukocyte cell-derived chemotaxin-2,LECT-2; hLECT2,LECT2
Detection method Sandwich Sensitivity 13.34 pg/mL
Standard 2000pg/mL Detection Range 31.25-2000pg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids
Reaction time 3.5H Research Area Signaling transduction;Cancer;Immunology;Metabolism
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human LECT2 .Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human LECT2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human LECT2 ,biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human LECT2 You can calculate the concentration of Human LECT2 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
2000.00 2.453 2.411
1000.00 1.631 1.589
500.00 1.042 1.000
250.00 0.626 0.584
125.00 0.259 0.217
62.50 0.220 0.178
31.25 0.138 0.096
0.00 0.042
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 53.931 334.857 810 82.483 212 747
Standard deviation 3.506 19.824 49.41 6.425 12.572 37.425
C V (%) 6.5 5.92 6.1 7.79 5.93 5.01
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 94-101 97
EDTA plasma (n=8) 90 93
Cell culture media (n=8) 93 95
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 91-100 88-97 94-100
Average (%) 95 92 97
1:4 Range (%) 94-103 89-99 89-97
Average (%) 99 94 93
1:8 Range (%) 89-97 95-102 89-96
Average (%) 93 99 93
1:16 Range (%) 98-107 90-96 92-101
Average (%) 102 93 97