Catalog Number:RE10072
TG(Triglyceride)ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name TG(Triglyceride)ELISA Kit Species Universal
Uniprot ID N/A Alternative Names TG
Detection method Competitive Sensitivity 1.85 µg/mL
Standard 200µg/mL Detection Range 3.13-200µg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids, Sample volume: 50μL
Reaction time 1.5H Research Area Metabolic pathway; Signal Transduction; Cancer, Nutrition metabolism;Hormone metabolism; Signal transduction;Metabolic pathway; Enzyme & Kinase; Signal transduction;Metabolic pathway; Nutrition metabolism;Hormone metabolism;Cardiovascular,Signaling transduction,Neuroscience,Cancer,Metabolism; Enzyme & Kinase;Infection immunity;Endocrinology;Hormone metabolism; Metabolic pathway; Signal Transduction; Cancer
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with TG. During the reaction, TG in the sample or standard competes with a fixed amount of TG on the solid phase supporter for sites on the Biotinylated Detection Ab specific to TG. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of TG in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(µg/mL) OD
200.00 0.178
100.00 0.343
50.00 0.699
25.00 0.873
12.50 1.262
6.25 1.536
3.13 2.076
0.00 2.775
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(µg/mL) 5.862 30.229 125.500 6.407 33.543 119.600
Standard deviation 0.290 2.297 5.585 0.299 1.228 7.738
C V (%) 4.95 7.60 4.45 4.66 3.66 6.47
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 96-104 100
EDTA plasma (n=8) 90-102 90
Cell culture media (n=8) 62-99 98
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 91-97 92-98 94-104
Average (%) 94 95 99
1:4 Range (%) 91-98 88-97 94-100
Average (%) 95 93 97
1:8 Range (%) 92-99 94-100 87-96
Average (%) 95 97 92
1:16 Range (%) 92-100 94-103 96-104
Average (%) 95 98 100
Assay Procedures