PRODUCTS

Catalog Number:RE5519H
Human ATP5F1B(ATP synthase F(1) complex subunit beta, mitochondrial) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name Human ATP5F1B(ATP synthase F(1) complex subunit beta, mitochondrial) ELISA Kit Species Human
Uniprot ID P06576 Alternative Names ATP synthase F(1) complex subunit beta, mitochondrial, ATP synthase F1 subunit beta, ATP5F1B, ATP5B, ATPMB, ATPSB
Detection method Sandwich Sensitivity 0.52 ng/mL
Standard 100ng/mL Detection Range 1.56-100ng/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids;Sample Volume=100μL
Reaction time 3.5H Research Area Cytokine;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human ATP5F1B .Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human ATP5F1B and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ATP5F1B ,biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human ATP5F1B You can calculate the concentration of Human ATP5F1B in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
100.00 2.325 2.276
50.00 1.770 1.721
25.00 1.253 1.204
12.50 0.579 0.530
6.25 0.286 0.237
3.13 0.219 0.170
1.56 0.127 0.078
0.00 0.049
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 3.886 14.243 41.700 2.552 10.871 58.600
Standard deviation 0.288 0.920 3.323 0.176 0.439 3.293
C V (%) 7.40 6.46 7.97 6.91 4.04 5.62
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 96-104 100
EDTA plasma (n=8) 88-95 92
Cell culture media (n=8) 95-102 98
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 92-101 92-100 88-98
Average (%) 96 96 93
1:4 Range (%) 88-96 96-104 97-105
Average (%) 92 101 102
1:8 Range (%) 90-99 96-102 91-99
Average (%) 95 99 94
1:16 Range (%) 94-103 91-100 93-100
Average (%) 99 96 97