PRODUCTS

Catalog Number:REA1900HF
MF-Human SOST(Sclerostin) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name MF-Human SOST(Sclerostin) ELISA Kit Species Human
Uniprot ID Q9BQB4 Alternative Names CDD, VBCH
Detection method Sandwich Sensitivity 4.38 pg/mL
Standard 800pg/mL Detection Range 12.5-800pg/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:25μL
Reaction time 2.5H Research Area N/A
Test principle N/A
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
800.00 2.479 2.439
400.00 1.839 1.799
200.00 1.333 1.293
100.00 0.697 0.657
50.00 0.364 0.324
25.00 0.154 0.114
12.50 0.140 0.100
0.00 0.040
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 31.145 141.486 500.800 19.586 107.314 339.600
Standard deviation 2.003 10.668 22.686 1.461 5.731 19.663
C V (%) 6.43 7.54 4.53 7.46 5.34 5.79
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 95-103 100
EDTA plasma (n=8) 95-105 100
Cell culture media (n=8) 95-105 100
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 89-96 97-106 97-103
Average (%) 92 101 100
1:4 Range (%) 96-104 97-105 95-103
Average (%) 101 100 100
1:8 Range (%) 89-97 94-100 94-102
Average (%) 92 97 99
1:16 Range (%) 95-103 88-96 97-107
Average (%) 99 93 102