PRODUCTS

Catalog Number:FSA1094R
FS-Rat TIMP-1(Tissue Inhibitors of Metalloproteinase 1) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Rat TIMP-1(Tissue Inhibitors of Metalloproteinase 1) ELISA Kit Species Rat
Uniprot ID P30120 Alternative Names EPO; CLGI; EPA; HCI; Erythroid Potentiating Activity; Collagenase Inhibitor; Metalloproteinase inhibitor 1; Fibroblast collagenase inhibitor
Detection method Sandwich Sensitivity 8.24 pg/mL
Standard 2000pg/mL Detection Range 31.25-2000pg/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:50μL
Reaction time 1.5H Research Area N/A
Test principle N/A
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
2000.00 2.349 2.306
1000.00 1.644 1.601
500.00 1.072 1.029
250.00 0.560 0.517
125.00 0.284 0.241
62.50 0.194 0.151
31.25 0.135 0.092
0.00 0.043
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 63.586 354.571 1063.000 84.897 304.286 1246.000
Standard deviation 3.052 15.495 40.819 3.727 13.875 59.933
C V (%) 4.80 4.37 3.84 4.39 4.56 4.81
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 91-99 95
EDTA plasma (n=8) 95-103 98
Cell culture media (n=8) 93-99 96
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 95-103 99-105 90-99
Average (%) 98 102 95
1:4 Range (%) 92-100 96-104 94-103
Average (%) 96 100 99
1:8 Range (%) 92-99 98-105 96-105
Average (%) 96 102 101
1:16 Range (%) 88-96 90-98 91-99
Average (%) 92 94 95