PRODUCTS

Catalog Number:REA3133HF
MF-Human LpPLA2(Lipoprotein-associated Phospholipase A2) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name MF-Human LpPLA2(Lipoprotein-associated Phospholipase A2) ELISA Kit Species Human
Uniprot ID N/A Alternative Names PLA2G7, LDL-PLA2, LP-PLA2, PAFAD, PAFAH, Lipoprotein-associated phospholipase A2, phospholipase A2 group VII
Detection method Sandwich Sensitivity 56.26 pg/mL
Standard 5000g/mL Detection Range 78.13-5000g/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:25μL
Reaction time 2.5H Research Area N/A
Test principle N/A
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
5000.00 2.437 2.362
2500.00 1.702 1.627
1250.00 1.024 0.949
625.00 0.692 0.617
312.50 0.346 0.271
156.25 0.194 0.119
78.13 0.120 0.045
0.00 0.075
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 128.621 880.000 2855.000 176.034 742.143 2337.500
Standard deviation 5.724 27.896 151.030 8.943 23.971 143.523
C V (%) 4.45 3.17 5.29 5.08 3.23 6.14
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 95-101 98
EDTA plasma (n=8) 95-105 100
Cell culture media (n=8) 91-99 96
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 95-104 94-103 89-99
Average (%) 99 98 94
1:4 Range (%) 97-105 89-95 90-98
Average (%) 100 92 93
1:8 Range (%) 92-99 96-105 88-97
Average (%) 96 100 93
1:16 Range (%) 96-103 88-96 92-101
Average (%) 100 93 97