PRODUCTS

Catalog Number:FS2873H
FS-Human IgG(Immunoglobulin G) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human IgG(Immunoglobulin G) ELISA Kit Species Human
Uniprot ID P01860(IGHG3) Alternative Names IgG, Immunoglobulin G, ImmunoglobulinG
Detection method Sandwich Sensitivity 1.45 ng/mL
Standard 200ng/mL Detection Range 3.13-200ng/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Signal transduction;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human IgG. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human IgG are added to the micro ELISA plate wells. Human IgG in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human IgG. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human IgG. The concentration of Human IgG in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
200.00 2.286 2.285
100.00 1.202 1.201
50.00 0.534 0.533
25.00 0.241 0.240
12.50 0.101 0.100
6.25 0.046 0.045
3.13 0.021 0.020
0.00 0.001 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 84.768 187.556 186.580 113.686 32.271 91.706
Standard deviation 11.697 3.457 9.284 12.110 4.385 3.107
C V (%) 9.470 0.078 5.249 7.750 0.351 3.527
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 90-100 92
EDTA plasma (n=8) 97-108 104
Cell culture media (n=8) 89-93 89
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-106 84-112 97-118
Average (%) 94 108 116
1:4 Range (%) 85-115 100-109 95-118
Average (%) 97 106 109
1:8 Range (%) 86-113 99-105 103-114
Average (%) 93 100 113
1:16 Range (%) 85-110 93-113 100-109
Average (%) 94 105 102
Assay Procedures